Leaf Optical Measurement Procedure - Standard Leaf

Standard Measurement Procedure for measuring both reflected and transmitted light on both sides of a leaf using an integrating sphere are:

View Measurement Orientation
Stray light No sample in sample port; Light source in reflected port.
Stray light reference No sample in sample port; Light source in reference port.
Stray light wall reference No sample in sample port; Light source in transmitted port.
Reflected top Sample in sample port with top of the sample facing into the sphere. Light source in reflected port.
Reference Top Sample in sample port with top of the sample facing into the sphere. Light source in reference port.
Transmitted Bottom Sample in sample port with top of the sample facing into the sphere. Light source in transmitted port.
Transmitted Top Sample in sample port with bottom of the sample facing into the sphere. Light source in transmitted port.
Reference Bottom Sample in sample port with bottom of the sample facing into the sphere. Light source in reference port.
Reflected Top Sample in sample port with top of the sample facing into the sphere. Light source in reflected port.